Skip to main content Skip to search Skip to main navigation

Frequently asked questions

Answers to the questions our customers ask most often – on ordering, design, documents and explosion protection.

Ordering and delivery

Do you make single units?
Yes. We manufacture from 1 piece, including ATEX versions.

What is the delivery time?
Standard versions approx. 2 weeks, custom builds on request. If it is urgent or you need a partial delivery, please say so in your enquiry.

Are there quantity prices?
Please state the quantities in your enquiry and we will take them into account in our quotation.

Do you build to drawing or sample?
Yes. Send us a drawing, sample or photo of the type plate by e-mail to info@ghi-gmbh.de – single units included.

Can you replace an existing sensor, including Schramm?
Yes. The number from the type plate, an old delivery note or a drawing is enough. We hold the order and production numbers of Schramm GmbH. See Schramm conversion.

Design

Can the sensor be shorter, longer or thinner than in the catalogue?
We make special lengths to your specification. Sheathed thermocouples are available from Ø 0.5 to 8 mm. Tell us the dimensions and installation and we will check the design.

Can a sheathed sensor be bent?
Yes, the sheath can be bent on site, as a guide with a radius of at least five times the diameter. The tip area containing the measuring element stays straight.

Which material are sheath and thermowell made of?
Standard sheath material is 1.4571 / 316Ti, above +600 °C Inconel 600 / 2.4816. Other materials on request. Thermowell materials are listed under Materials, the version of each item in its data sheet.

Which protection class does the transition sleeve have?
The standard transition sleeve is crimped. For humid environments it is available potted to IP67, for permanent immersion to IP68.

Which connecting cables are available?
Silicone up to 180 °C, PTFE/PFA up to 260 °C, glass fibre with stainless steel braid up to 400 °C. Thermocouples use thermocouple or compensating cable matching the type, Pt100 sensors copper cable with 2, 3 or 4 wires per element. The configurator of each series shows the exact construction.

Is higher accuracy or another thermocouple type possible?
Pt100 and Pt1000 in class B, A or AA to IEC 60751; the validity range depends on element and diameter and is stated in the quotation. Thermocouples type K, N, J and S in class 1 or 2 to IEC 60584-1.

Documents and testing

Is there a data sheet or dimensional drawing?
Yes. Data sheets with dimensional drawing and order code are in the download centre; the configurator lets you put together your version.

Can I get a RoHS declaration?
Yes, a RoHS declaration is available for all sensors on request.

Is an inspection certificate 3.1 or a calibration available?
On request we supply an inspection certificate 3.1 to EN 10204 and a factory calibration with factory calibration certificate. DAkkS calibration and calibration to CQI-9 are offered through an accredited partner laboratory.

Up to which pressure is a thermowell suitable?
That depends on material, wall thickness, temperature and flow. We calculate your thermowell to ASME PTC 19.3 TW-2016 and send you the report, also for thermowells already installed. See Thermowell calculation.

Explosion protection

Is there an Ex version, and what is the marking?
Yes, for the measuring inserts WME-EX and TME-EX and the sheathed sensors WML-EX and TML-EX, certified under EU type examination certificate EPS 25 ATEX 1 060 X Rev. 1 and IECEx EPS 25.0012X:

WME-EX, WML-EX, TME-EX, TML-EX without protective tube
EPS 25 ATEX 1 060 X Rev. 1 · IECEx EPS 25.0012X
II 1G Ex ia IIC T1…T6 Ga
II 2G Ex ib IIC T1…T6 Gb
II 2D Ex ib IIIC T440 °C…T63 °C Db
Sensor tip −50…+450 °C · connection head −40…+85 °C
Measuring insert WME-EX or TME-EX in a protective tube
EPS 25 ATEX 1 060 X Rev. 1 · IECEx EPS 25.0012X
II 1/2G Ex ib IIC T1…T6 Ga/Gb
II 1/2D Ex ib IIIC T200 440 °C…T200 63 °C Da/Db
Sensor tip −50…+450 °C · connection head −40…+85 °C

For other designs in Ex version, please enquire. Zones and marking are explained on Explosion protection ATEX.

Your question is not listed? Send an enquiry → or +49 6181 5049318 · info@ghi-gmbh.de