Thermocouples: types and tolerances to DIN EN 60584

A thermocouple generates a temperature-dependent thermoelectric voltage at the junction of two metals — the Seebeck effect. It is robust, fast and can be used up to 1,700 °C.

Which type suits which purpose, how large the permissible deviation is and which voltage is present at which temperature is set out in DIN EN 60584-1.

The types at a glance

TypeMaterial pairing (+ / −)Application rangeProperty
KNiCr – NiAl−200…+1,200 °Cstandard type, oxidation-resistant
NNiCrSi – NiSi−200…+1,300 °Cmore stable than K in long-term use
JFe – CuNi−40…+750 °Ceconomical, reducing atmosphere
TCu – CuNi−200…+350 °Clow temperatures, refrigeration
ENiCr – CuNi−200…+900 °Chighest thermoelectric voltage
SPt10Rh – Pt0…+1,600 °Cnoble metal, glass and calibration
RPt13Rh – Pt0…+1,600 °Cnoble metal, like S
BPt30Rh – Pt6Rh+600…+1,700 °Chighest temperatures

Colour coding of cables to IEC 60584-3: sheath and positive conductor in the type colour, negative conductor white.

Colour coding to IEC 60584-3

Sheath and positive conductor carry the type colour, the negative conductor is always white. This lets you identify thermocouple and compensating cables without a measuring instrument.

TypeSheath and positive conductorNegative conductorColour
Kgreenwhite#009640
Jblackwhite#1A1A1A
Npinkwhite#F2A0BD
Tbrownwhite#7B4B2A
Evioletwhite#6F3F9E
R, Sorangewhite#EF7D00
Bgreywhite#8C8C8C

Colours to IEC 60584-3. The marking on the supplied cable remains authoritative.

Tolerances to IEC 60584-1

TypeClass 1Class 2Class 3
T−40…350: ± 0.5 °C or 0.004·|t|−40…350: ± 1.0 °C or 0.0075·|t|−200…40: ± 1.0 °C or 0.015·|t|
E−40…800: ± 1.5 °C or 0.004·|t|−40…900: ± 2.5 °C or 0.0075·|t|−200…40: ± 2.5 °C or 0.015·|t|
J−40…750: ± 1.5 °C or 0.004·|t|−40…750: ± 2.5 °C or 0.0075·|t|–
K, N−40…1,000: ± 1.5 °C or 0.004·|t|−40…1,200: ± 2.5 °C or 0.0075·|t|−200…40: ± 2.5 °C or 0.015·|t|
R, S0…1,100: ± 1.0 °C; 1,100…1,600: ± [1 + 0.003·(t − 1,100)]0…1,600: ± 1.5 °C or 0.0025·|t|–
B–600…1,700: ± 0.0025·|t|600…1,700: ± 4.0 °C or 0.005·|t|

Reference temperature 0 °C. The larger of the two values applies.

Thermoelectric voltage in mV, reference junction 0 °C

°CType KType NType JType S
−200−5.891–−7.890–
−100−3.554−2.407−4.633–
00000
1004.0962.7745.2690.646
2008.1385.91310.7791.441
30012.2099.34116.3272.323
40016.39712.97421.8483.259
50020.64416.74827.3934.233
60024.90520.61333.1025.239
70029.12924.52739.1326.275
80033.27528.455–7.345
90037.32632.371–8.449
1,00041.27636.256–9.587
1,10045.11940.087–10.757
1,20048.83843.846–11.951

Characteristic curves of the thermoelectric voltage in millivolts over temperature for types K, N, J and S

Thermoelectric voltage over temperature, reference junction 0 °C. Type J rises most steeply but ends at +750 °C; type S gives the smallest voltage and reaches furthest.

Measuring junction insulated or grounded

  • Insulated — the thermocouple is galvanically isolated from the sheath. Interference-proof, our standard.
  • Grounded — the thermocouple is welded to the sheath bottom. Shorter response time, half-value time from about 0.15 s for small diameters.

Longitudinal section through the sheath tip: left, thermocouple isolated from the sheath; right, welded to the sheath bottom

Insulated or grounded — the difference lies in the tip and determines interference immunity and response time.

Sheath materials

Inconel 600 / 2.4816 up to 1,150 °C, 1.4571 and 1.4404 up to 800 °C. Special materials such as Nicrofer, Hastelloy, Inconel 601 and tantalum on request. The complete list of materials can be found under Materials.

Which type for your measuring point?

Tell us medium, temperature and required accuracy — we will tell you which type and class make sense there.